ImageCoating thickness gauges TM 510 FN is professional level gauges for industrial users who need precise and easy to use measurements with statistic functions.

 

  • Measuring modes: continuous/single
  • Shutdown modes: manual/automatic
  • Wide measuring range with 5 probes available (next page)
  • Direct testing mode and block statistics mode (APPL/BATCH)
  • Direct print out of statistical values
  • Dataview to connect with PC
  • 500 data can be stored

Technical specifications

Image 

 

Image

 

Standard delivery

 

Main unit

1

Probe F1 or N1

1

Charger

1

Calibration foils

1

Instruction manual

1

TMTeck cetificate

1

Warranty card

1

Carrying case

1

 

Optional accessories

  • 5 probes for different applications
  • PC software Dataview for online and data transfer
  • Calibration foils in various thickness
  • Communication cable

 

Image

Option probes for TM510FN

 

Type F probe

Probe

F400

F1

F1/90°

F10

Measuring Principle

Magnetic method

Measuring range(mm)

0400

01250

010000

Min resolution(mm)

0.1

0.1

10

Tolerance

One-point calibration (mm)

±(2%H+0.7)

±(2%H+1)

±(2%H+10)

Two-point calibration (mm)

±(1%H+0.7)

±((1%H+1)

±(1%H+10)

Minimum radius of curvature (convex) (mm)

1

1.5

flat

10

 Minimum measuring area(mm)

F3

F7

F7

F40

Minimum thickness of base material(mm)

0.2

0.5

0.5

2

             

 

Type N probe

Probe

N400

N1

CN02

Measuring Principle

Eddy current method

 Measuring range (mm)

0400

01250

10200

 Min resolution(mm)

0.1

0.1

1

Tolerance

 One-point calibration (mm)

±(2%H+0.7)

±(2%H+1.5)

±(2%H+1)

 Two-point calibration (mm)

±(1%H+0.7)

±1%H+1.5

--------

Minimum radius of curvature (convex) (mm)

1.5

3

flat

Minimum measuring area(mm)

F4

F5

F7

Minimum thickness of base material(mm)

0.3

0.3

0